Natural-Knee¿ II Posterior Stabilized (P.S.) Condylar Tibial Inserts: N-K II CONS PS TIB INS SZ N-K II CONS PS TIB INS SZ N-K II CONS PS TIB INS SZ N-K II CONS PS TIB INS SZ N-K II CONS PS TIB INS SZ N-K II CONS PS TIB INS SZ N-K II CONS PS TIB INS SZ N-K II CONS PS INS TIB SZ N-K II CONS PS TIB INS SZ N-K II CONS PS TIB INS SZ N-K II CONS PS TIB INS SZ N-K II CONS PS TIB INS SZ N-K II CONS PS TIB INS SZ N-K II CONS PS TIB INS SZ

Z-1257-2014

February 25, 2014

Class II

Product summary

Firm
Zimmer, Inc.
Event
Event 67503
Status
Terminated
Classification
Class II
Quantity
3638 units
Official record key
device-enforcement:Z-1257-2014

Official wording

Reason: A complaint was received for a discolored All Poly Tibia. Root cause analysis found that the discoloration was the result of free radical oxidation. The oxidation process was advanced prematurely as a result of the machined component being stored in a transparent container exposed to artificial light for six years prior to inert packaging and sterilization of the device.

Code information: Item Number 672509100 672509101 672509200 672509203 672511100 672511103 672513101 672513103 672513201 672516103 672519103 672522201 672525103 672525203

Distribution pattern: Worldwide Distribution-USA (nationwide) including DC and the states of AK, AL, AR, AZ, CA, CO, CT, FL, IL, IN, KS, LA, ME, MI, MO, MS, NC, ND, NE, NJ, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, and WI, and the countries of Canada, Australia, India, Austria, Japan, Czech Republic, Seoul Korea, France, Venezuela, Germany, Luxembourg, Spain, and Switzerland.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    A complaint was received for a discolored All Poly Tibia. Root cause analysis found that the discoloration was the result of free radical oxidation. The oxidation process was advanced prematurely as a result of the machined component being stored in a transparent container exposed to artificial light for six years prior to inert packaging and sterilization of the device.