Device product
AVEA Standard ventilator-refurbished, R17311-xx (xx = 0-14) A continuous ventilator intended to mechanically control or assist broad spectrum patient breathing by delivering a predetermined percentage of oxygen in the breathing gas.
Z-1892-2016
Product summary
- Event
- Event 74111
- Status
- Terminated
- Classification
- Class I
- Quantity
- 922 units total (501 units in US)
- Official record key
device-enforcement:Z-1892-2016
Official wording
Reason: CareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to the User Interface Module (UIM).
Code information: Unavailable
Distribution pattern: Worldwide Distribution - US (Nationwide) and the countries of Australia, Hungary, Paraguay, Belgium, Indonesia, Philippines, Bolivia, Ireland, Poland, Bermuda, India, Saudi Arabia, Canada, Japan, Singapore, Chile, Jordon, South Africa, China, Kuwait, Taiwan, Ecuador, Malaysia, Thailand, Egypt, Mexico, Turkey, France, Netherlands, United Kingdom, Guatemala, and Nicaragua.
Derived failure modes
-
Unknown
CareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to the User Interface Module (UIM).