Device product

TSX-101A Aquilion 32/64 ONE Computed Tomography X-Ray System; TXS-101/R Aquilion RXL Computed Tomography X-Ray System; TSX-301A Aquilion One Computed Tomography X-Ray System; TSX-301B Aquilion Premium Computed Tomography X-Ray System; TSX-302A Aquilion Prime Computed Tomography X-Ray System. This device is indicated to acquire and display cross sectional volumes of the whole body, to include the head. The Aquilion RXL has the capability to provide volume sets. These volume sets can be used to perform specialized studies, using indicated software/hardware, by a trained and qualified physician.

Z-1113-2014

October 01, 2013

Class II

Product summary

Firm
Toshiba American Medical Systems Inc
Event
Event 66527
Status
Terminated
Classification
Class II
Quantity
338 units
Official record key
device-enforcement:Z-1113-2014

Official wording

Reason: Toshiba intitiated this recall because their investigation revealed that due to a problem of the X-ray output control software, the X-ray output conditions may not be set correctly. As a result, X-ray exposure under the scan conditions used for contrast study may not start.

Code information: SERIAL NUMBER JDA1062176, JDA1062177, JDA1072183, JDA1072184, JDA1072186, JDA1072187, JDA1082188, JDA1082189, JDA1082190, JDA10X2191, JDA10X2192, JDA10X2193, JDA10X2194, JDA10X2195, JDA10X2196, JDA10X2197, JDA10X2198, JDA10Y2199, JDA10Y2200, JDA10Y2201, JDA10Y2202, JDA10Y2203, JDA10Y2204, JDA10Y2205, JDA10Z2206, JDA10Z2207, JDA10Z2209, JDA10Z2210, JDA10Z2211, JDA10Z2212, JDA10Z2213, JDA10Z2214, JDA10Z2215, JDA1112216, JDA1112217, JDA1112218, JDA1112219, JDA1112221, JDA1112222, JDA1112224, JDA1112225, JDA1112226, JDA1112227, JDA1122231, JDA1122234, JDA1142236, JDA1162237, JDA1162238, JDA1162239, JDA1182240, JDA1182241, JDA1182242, JDA1182243, JDA1182244, JDA11X2245, JDA11X2246, JDA11X2247, JDA11Z2249, JDA11Z2250, JDA11Z2252, JDA11Z2253, JDA11Z2254, JDA11Z2255, JDA1222257, JDA1222258, JDA1222259, JDA1222260, JDA1222261, JDA1242264, JDA1252265, JDA1252266, JDA1252267, JDA12Z2271, JDA1312272, JGA1062178, JGA1062181, JGA1062182, JGA1112220, JGA1112228, JGA11Z2256, JGA1222262, JGA12X2269, JLA1062179, JLA1062180, JLA1072185, JLA10Z2208, JLA1122232, JLA1122233, JLA1132235, JLA11Y2248, JLA11Z2251, JLA1232263, JLA1262268, JLA12Y2270, KDA1062080, KDA1072081, KDA1072082, KDA1072083, KDA1072085, KDA1072086, KDA1072088, KDA1082090, KDA1082091, KDA1082092, KDA1092093, KDA10X2094, KDA10X2095, KDA10X2096, KDA10X2097, KDA10X2098, KDA10X2099, KDA10Y2100, KDA10Y2101, KDA10Y2102, KDA10Z2104, KDA1112107, KDA1122110, KDA1122111, KDA1132113, KDA1142114, KDA1152115, KDA1152116, KDA1152117, KDA1152118, KDA1162119, KDA1162120, KDA1162121, KDA1162122, KDA1162123, KDA1162124, KDA1172125, KDA1182126, KDA1182127, KDA1182128, KDA1182129, KDA1192130, KDA1192131, KDA11X2132, KDA11X2133, KDA11Y2134, KDA11Z2136, KDA11Z2137, KDA11Z2138, KDA11Z2139, KDA11Z2140, KDA1222141, KDA1222142, KDA1222143, KDA1232144, KDA1242145, KDA1242146, KDA1252147, KDA1272148, KDA1282149, KDA1292151, KDA12X2152, KDA12Y2153, KDA12Z2154, KDA1322155, KGA1082089, KGA10Z2103, KGA10Z2105, KGA1112106, KGA1112108, KGA1122109, KGA1122112, KGA11Z2135, KLA1072084, KLA1072087, KLA1292150, RDA1292019, RDA12X2023, RDA12X2024, RDB12Y2029, RDB12Z2032, RDB1312037, RDB1312038, RGA12X2025, RGB12Y2028, RGB1312034, 2DA0792002, 2DA0792004, 2DA07X2005, 2DA07X2008, 2DA07Y2010, 2DA0842023, 2DA0842026, 2DA0862034, 2DA0862036, 2DA0872038, 2DA0882048, 2DA0882049, 2DA0882051, 2DA0882052, 2DA0882053, 2DA08X2061, 2DA08X2067, 2DA08X2070, 2DA08X2071, 2DA08X2072, 2DA08Y2077, 2DA08Y2080, 2DA08Z2092, 2DA0952115, 2DA0962116, 2DA0972123, 2DA0982131, 2DA0982133, 2DA09X2139, 2DA09X2143, 2DA09Y2148, 2DA09Y2151, 2DA09Z2158, 2DA09Z2161, 2DA09Z2162, 2DA1012165, 2DA1012167, 2DA1012168, 2DA1012169, 2DA1012170, 2DA1012174, 2DA1062208, 2DA1072212, 2DB1082219, 2DB1082220, 2DB1082222, 2DB10X2241, 2DB10Y2253, 2DB10Y2257, 2DB10Z2265, 2DB1112275, 2DB1112283, 2DB1122284, 2DB1122291, 2DB1122294, 2DC1142308, 2DC1142311, 2DC1152316, 2DC1162324, 2DC1172328, 2DC1172332, 2DC1172333, 2DC11X2352, 2DC11Y2365, 2DC11Y2367, 2DC11Z2372, 2DC11Z2374, 2DC1212377, 2DC1212383, 2DC1232392, 2DC1232393, 2DC1232396, 2DC1232397, 2DC1232398, 2DC1242407, 2DC1242409, 2DC1252410, 2DC1262412, 2DC1272417, 2DC1272421, 2DC1282423, 2DC12X2427, 2DC12Y2430, 2DC12Y2431, 2DC12Z2433, 2DC1312435, 2DC1312436, 2DC1322437, 1DA0962006, 1DA0972007, 1DA0972008, 1DA0992011, 1DA09Y2014, 1DA1062015, 1DB1072016, 1DB1082017, 1DB1082018, 1DB1082019, 1DB10Y2020, 1DB10Y2021, 1DB10Y2022, 1DB1112025, 1DB1122027, 1DC1142030, 1DC1152031, 1DC1162032, 1DC1162033, 1DC1172035, 1DC11Y2037, 1DC11Y2038, 1DC11Z2040, 1DC1222042, 1DC1232043, 1DC1232044, 1DC1252045, 1DC1252046, 1DC1262048, 1DC1272050, 1DC1282051, 1DC1292052, 1DC12X2053, 1DC12Y2054, 1DC12Y2055, 1DC12Z2056, 1DC1312057, 1DC1312058, 1DC1342059, 1GC1262047, 1GC1272049, 1DA1222096, 1DA1222097, 1DA1232099, 1DA1232100, 1DA1232103, 1DA1252124, 1DA1252131, 2DA1252001, 2DA1262003, 2DA1272004, 2DA1272005, 2DA1272006, 2DA1282007, 2DA1292008, 2DA12X2009, 2DA12X2010, 2DA12X2012, 2DA12X2013, 2DA12Y2014, 2DA12Y2015, 2DA12Y2016, 2DA12Z2017, 2DA12Z2018, 2DA1312019, 2DA1312020, 2DA1312021, 2DA1312022, 2GA1262002, 2GA12X2011, 2DA1072211, 2DA08X2063, 2DC11Y2368.

Distribution pattern: Nationwide Distribution.

Derived failure modes

  • Device software or design

    reason.device_software_design · v1.0.0

    problem of the X-ray output control software