Skip to content
Recall Observatory FDA recall evidence

Device product

Zilver PTX Drug-Eluting Peripheral Stent, - 7 mm X 80mm; 125 cm, Rx, Sterile Product Usage: The Zilver PTX Drug -Eluting Peripheral Stent is a self expanding, small metal, mesh tube (stent) with the outer surface coated with the drug Paclitaxel that can be Implanted in an artery in the thigh (femoropopliteal artery). The Paclitaxel coating helps prevent the artery from narrowing again (restenosis). The Zilver; PTX stent acts as a scaffold to: Hold open a narrowed artery in the thigh (femoropopliteal artery) caused by PAD. Improve blood flow to the extremity.

Z-1275-2013

April 18, 2013

Class I

Product summary

Firm
Cook, Inc.
Event
Event 64975
Status
Terminated
Classification
Class I
Quantity
535 units in USA
Official record key
device-enforcement:Z-1275-2013

Official wording

Reason: Cook Medical has received a small number of complaints related to the delivery system for the Zilver PTX Drug-Eluting Peripheral Stent involving fractures of the delivery system inner catheter after stent deployment, and separation of the inner catheter tip section (13 complaints with an occurrence rate of 0.043%) Two adverse events, including one death, occurred in cases where a tip separation

Code information: REF ZIV6-35-125-7-80-PTX ; UDI 10827002248950 all lots distributed through 4/16/13

Distribution pattern: Worldwide Distribution - USA Nationwide; in the countries of Japan, Europe, Brazil, and Australia

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Cook Medical has received a small number of complaints related to the delivery system for the Zilver PTX Drug-Eluting Peripheral Stent involving fractures of the delivery system inner catheter after stent deployment, and separation of the inner catheter tip section (13 complaints with an occurrence rate of 0.043%) Two adverse events, including one death, occurred in cases where a tip separation