Skip to content
Recall Observatory FDA recall evidence

Device product

ORTHOFIX, Connector System, LARGE SET SCREW, REF 79-2002

Z-1364-2018

February 23, 2018

Class II

Product summary

Firm
Orthofix, Inc
Event
Event 79498
Status
Terminated
Classification
Class II
Quantity
188 units
Official record key
device-enforcement:Z-1364-2018

Official wording

Reason: Orthofix is conducting a voluntary recall of the Connector System Large Set Screw (PN: 79-2002) with affected Lot Numbers 001 and 002, and Small Set Screw (PN: 79-2003) with affected Lot Numbers 002, 003 and 004. Through the company's manufacturing controls, size discrepancies were noted with the drive feature for both the large and small sets screws. As a result, it is possible that (1) the set screw drive feature will not accept the Set Screw Driver (PN: 79-1006) or (2) the driver may get wedged in the corners of the drive feature (false bottom) which may lead to a stripped set screw.

Code information: UDI 18257200125364, Lot Numbers: O01 and O02

Distribution pattern: US

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Orthofix is conducting a voluntary recall of the Connector System Large Set Screw (PN: 79-2002) with affected Lot Numbers 001 and 002, and Small Set Screw (PN: 79-2003) with affected Lot Numbers 002, 003 and 004. Through the company's manufacturing controls, size discrepancies were noted with the drive feature for both the large and small sets screws. As a result, it is possible that (1) the set screw drive feature will not accept the Set Screw Driver (PN: 79-1006) or (2) the driver may get wedged in the corners of the drive feature (false bottom) which may lead to a stripped set screw.