Device product
X-COATED SMALL PATIENT PACK Product Usage: The Sarns Cardioplegia Cannula and Cardioplegia/Vent Catheters are indicated for placement in the aortic root to deliver cardioplegia solution or blood to the heart, to vent the left heart, and to aspirate air from the aorta during cardiopulmonary bypass procedures.
Z-1883-2012
Product summary
- Event
- Event 61731
- Status
- Terminated
- Classification
- Class II
- Quantity
- 53,139 (each) in total
- Official record key
device-enforcement:Z-1883-2012
Official wording
Reason: During production of the Sarns Antegrade Cardioplegia Cannula, Terumo Cardiovascular Systems (Terumo CVS) identified the presence of a foreign substance on the inner surface of some cannulae tips. Terumo CVS' preliminary investigation found that the substance can be dislodged from the cannula surface. The substance was likely deposited during the molding process, but the exact composition of the
Code information: Catalog number: 71936-01 and lot number: MH06, ML20, MN15, NG20, and NG20A.
Distribution pattern: Worldwide Distribution - US (nationwide) including the states of: AL, AR , AZ , CA, CT , FL, IA , IL, IN , KS , LA , MI , NM, MO , NC, NE , NJ , NV, NY , OH, OK , PA, RI , SC , TN, TX, WA , WI, WV and the countries of: BELGIUM, BRAZIL, Byelorussian SSR, CAMBODIA , Dominican Republic, Honduras, Japan, Malaysia, Mexico, Nicaragua, Saudi Arabia, Singapore, South Africa, Thailand, Trinidad and Tobago, UNITED ARAB EMIRATES (UAE), and Vietnam.
Derived failure modes
-
Foreign material or chemical contamination
foreign substance