Skip to content
Recall Observatory FDA recall evidence

Device product

Hologic Selenia Full Field Digital Mammography System: SEL-00002, SEL-00013, SEL-00030, SEL-00039. The Selenia Full Field Digital Mammography system generates digital mammography images that can be used for screening and diagnosis of breast cancer.

Z-1700-2014

May 13, 2014

Class II

Product summary

Firm
Hologic, Inc.
Event
Event 68298
Status
Terminated
Classification
Class II
Quantity
11 units
Official record key
device-enforcement:Z-1700-2014

Official wording

Reason: Selenia Full Field Digital Mammography System defective relay switch, if there is a need to use the emergency off switch, the system may not turn off which could result in patient injury

Code information: Serial Numbers: 2840314W8065, 2840314W8062W, 2840314W8066W, 2840314W8067 2840314W8073, 2840314W8064, 2840314W8056W, 28401082720 28407083579, 28401137611W, 28403083045W

Distribution pattern: Distributed in Puerto Rico, and the states of CA, NY, and TX, and the country of Republic of Korea.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Selenia Full Field Digital Mammography System defective relay switch, if there is a need to use the emergency off switch, the system may not turn off which could result in patient injury