Skip to content
Recall Observatory FDA recall evidence

Device product

Siemens Mobilett XP Digital Mobile X-ray System Product Usage: The Mobilett XP and Mobilett XP Hybrid are radiographic systems designed for use in wards, intensive care, and premature birth-wards, pediatric and emergency departments, operating theatres as well as the central X-ray department.

Z-1985-2014

May 09, 2014

Class II

Product summary

Firm
Siemens Medical Solutions USA, Inc
Event
Event 68382
Status
Terminated
Classification
Class II
Quantity
10
Official record key
device-enforcement:Z-1985-2014

Official wording

Reason: There is a potential fire hazard of certain Lithium-Ion batteries in the computers used in the original manufacturing or as spare parts between October 2010 and April 2011 for the Siemens Mobilett XP Digital mobile X-ray System, which might occur at any time with higher probability while the system is being charged and connected to the main power supply.

Code information: model no. 1818447, with serial nos.: 3284 3449 1026 1025 1010 1041 1054 2002 2058 2022

Distribution pattern: US Nationwide Distributions in the states of MO, GA, FL, NJ, CA, PA and including NM.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    There is a potential fire hazard of certain Lithium-Ion batteries in the computers used in the original manufacturing or as spare parts between October 2010 and April 2011 for the Siemens Mobilett XP Digital mobile X-ray System, which might occur at any time with higher probability while the system is being charged and connected to the main power supply.