115 V Blanketrol II, Model 222S Hyper-Hypothermia System . Used to lower or to raise a patients temperature and /or maintain a desired patient temperature through conductive heat transfer. Part number 86165.

Z-1679-2014

January 08, 2014

Class II

Product summary

Firm
Cincinnati Sub-Zero Products Inc
Event
Event 68356
Status
Terminated
Classification
Class II
Quantity
7 units
Official record key
device-enforcement:Z-1679-2014

Official wording

Reason: The incorrect voltage component was placed into the device during manufacturing. This may make the device unable to maintain patient temperature during cooling mode. The device will continue to indicate that the unit is cooling because the compressor will continue to run, however water in the reservoir will begin to return to ambient temperature. The actual water temperature will be displayed but

Code information: Model Number 222S; Serial #s: 133-2-00697, 133-2-00698; 133-2-00699; 133-2-00700; 133-2-00701; 133-2-00702; and 133-2-00703:.Part # 86165

Distribution pattern: The product was distributed to one customer in MI, and internationally (Taiwan)

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The incorrect voltage component was placed into the device during manufacturing. This may make the device unable to maintain patient temperature during cooling mode. The device will continue to indicate that the unit is cooling because the compressor will continue to run, however water in the reservoir will begin to return to ambient temperature. The actual water temperature will be displayed but