Device product
Tri-Staple 2.0 Curved Tip Intelligent Reload for use with Signia stapling system 45 mm Vascular/Medium, Item Code SIG45ACTAVM
Z-2515-2018
Product summary
- Event
- Event 80423
- Status
- Ongoing
- Classification
- Class II
- Quantity
- Unavailable
- Official record key
device-enforcement:Z-2515-2018
Official wording
Reason: Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
Code information: N7L1049X & N7M0723X
Distribution pattern: Nationwide U.S.A. Internationally to: Albania, Algeria, Austria, Belgium, Bulgaria, Croatia, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Guadeloupe, Hungary, Iceland, Ireland, Israel, Italy, Ivory Coast, Jordan, Kazakhstan, Kuwait, Latvia, Lebanon, Luxembourg, Macedonia, Malta, Martinique, Moldova, Morocco, Namibia, Netherlands, New Caledonia, Norway, Pakistan, Poland, Portugal, Reunion, Romania, Russian federation, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Tunisia, Turkey, United Arab Emirates & United Kingdom
Derived failure modes
-
Unknown
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.