Device product
Part # 313.212, 1.3mm Stardrive Screwdriver Bld Slf-Retain Med/60mm MQC, Lot # 4363780, mfg. 9/18/2002 Product Usage: The DePuy Synthes MR Safe Instruments and MR conditional Instruments are used to assist with the insertion or removal of an implant.
Z-2195-2016
Product summary
- Event
- Event 74416
- Status
- Terminated
- Classification
- Class II
- Quantity
- 34 units
- Official record key
device-enforcement:Z-2195-2016
Official wording
Reason: DePuy Synthes is voluntarily initiating this recall due to affected devices that were distributed, etched and or labeled with MR Safety information not meeting the current ASTM standard.
Code information: Part # 313.212, 1.3mm Stardrive Screwdriver Bld Slf-Retain Med/60mm MQC, Lot # 4363780, mfg. 9/18/2002
Distribution pattern: US Nationwide Distribution in the states of: AR,CA,CO,CT,DC,ID,IL,KY,LA,MA,MD,MN,MO,MS,MT,NY,PA and TX.
Derived failure modes
-
Unknown
DePuy Synthes is voluntarily initiating this recall due to affected devices that were distributed, etched and or labeled with MR Safety information not meeting the current ASTM standard.