Device product

Hitachi Oasis MRI system Product Usage: Hitachi MR system is an imaging device and is intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation. The Oasis system produces transverse, coronal, sagittal, oblique, and curved crosssectional images that display the internal structure of the head, body, or extremities. The images produced by the MR systems reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2), and flow. When interpreted by a trained physician, these images provide information that can be useful in diagnosis determination.

Z-2479-2018

June 18, 2018

Class II

Product summary

Firm
Hitachi Medical Systems America Inc
Event
Event 80464
Status
Terminated
Classification
Class II
Quantity
264
Official record key
device-enforcement:Z-2479-2018

Official wording

Reason: The stainless steel belt which runs the length of the table underneath the tabletop is exposed when the tabletop is extended into the magnet for scanning. The firm has received two complaints where technologists have cut their fingers on this belt.

Code information: M001 to M028, M030 to M108, M110 to M266, M951, M952

Distribution pattern: Worldwide Distribution - US Nationwide distribution.in the states of Mexico.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The stainless steel belt which runs the length of the table underneath the tabletop is exposed when the tabletop is extended into the magnet for scanning. The firm has received two complaints where technologists have cut their fingers on this belt.