RSD 800 Sorting System calculator/data processing module, for clinical use Product Usage: The cobas 8000 Modular Analyze Series is a fully automated system for Clinical chemistry analysis, intended for the in vitro qualitative and quantitative determination of analytes in body fluids. It is optimized for high throughput workloads in the professional environment using a combination of ion selective electrodes (ISE) and a photometric analysis unit.

Z-0105-2017

July 15, 2016

Class II

Product summary

Firm
Roche Diagnostics Operations, Inc.
Event
Event 75186
Status
Terminated
Classification
Class II
Quantity
9014 in total
Official record key
device-enforcement:Z-0105-2017

Official wording

Reason: Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers sample rack tray. The cleaning procedure for the sample rack tray has been revised to clarify the cleaning process.

Code information: Not applicable

Distribution pattern: US Nationwide Distribution

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers sample rack tray. The cleaning procedure for the sample rack tray has been revised to clarify the cleaning process.