Skip to content
Recall Observatory FDA recall evidence

Device product

Stryker brand Arthroscope and Hardware Set Tray (sterilization tray); Model/Part Number: 233-032-116; Product is manufactured and distributed by Stryker Endoscopy, San Jose, CA The Arthroscope and Hardware Set Tray is used to hold and protect surgical devices during the sterilization process. Specifically this tray is used for arthroscopes and arthroscopic hardware devices. The tray consists of an interlocking tray and lid which are perforated to allow for the passage of sterilizing agent from outside the tray to the inside of the tray where the devices are held.

Z-1961-2012

June 19, 2012

Class II

Product summary

Firm
Stryker Endoscopy
Event
Event 62346
Status
Terminated
Classification
Class II
Quantity
6388 units
Official record key
device-enforcement:Z-1961-2012

Official wording

Reason: The gravity steam, Ethylene Oxide (EO) and STERRAD 100S sterilization methods are being removed from the "Instructions for Use"; due to the firm's evaluation that identified original validation was not done at worst case conditions.

Code information: All serial/lot numbers

Distribution pattern: Worldwide Distribution--USA (nationwide)

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The gravity steam, Ethylene Oxide (EO) and STERRAD 100S sterilization methods are being removed from the "Instructions for Use"; due to the firm's evaluation that identified original validation was not done at worst case conditions.