Alere Triage Cardiac Panel PN 97000HS Used for the quantitative determination of creatine kinase MD (CK--MB), myoglobin and troponin I in EDTA anticoagulated whole blood or plasma specimens.
Z-1929-2012
Product summary
- Event
- Event 61927
- Status
- Terminated
- Classification
- Class I
- Quantity
- 6,044 kits
- Official record key
device-enforcement:Z-1929-2012
Official wording
Reason: Certain lots of the affected products may have significantly decreased precision relative to the package insert, which could result in an increased frequency of false positive or false negative results. These false positive and false negative results are unpredictable within lots and may not be detected by Quality Control testing.
Code information: Phase 1: K50514B, W49753B, W49754B, W49756B, W49757B, W49758B, W49760B, W49761B, W49762B, W49764B, W49767B, W49768B, W49769B, W49770B, W50608B, W50613B, W50621B, W50623B, W50625B, W50632B, W50634B, W50637B, W50639B, W50641B, W50642B, W51114B, and W51119B. Phase 2: W49766B, W50605B, W50627B, W50628B, and W50629B.
Distribution pattern: Nationwide Distribution
Derived failure modes
-
Unknown
Certain lots of the affected products may have significantly decreased precision relative to the package insert, which could result in an increased frequency of false positive or false negative results. These false positive and false negative results are unpredictable within lots and may not be detected by Quality Control testing.