PRISM MEDICAL A-625, MAX. LOAD 625 lbs. 283 Kgs., Patient lift, non-A/C powered

Z-0013-2018

September 08, 2017

Class II

Product summary

Firm
Handicare Usa Inc
Event
Event 78096
Status
Terminated
Classification
Class II
Quantity
8 units
Official record key
device-enforcement:Z-0013-2018

Official wording

Reason: During internal cycle testing of A-Series lifts Handicare has found that the A-Series lift strap for lifts manufactured between April 25, 2017 and June 12, 2017 may wear prematurely at max load (625 lbs). We have identified the root cause as a supplied component that was not to specification. If the strap wears during use there is a potential risk for the person being supported by the lift to drop.

Code information: A6M0001526, A6M0001527, A6M0001528, A6M0001529, A6M0001530, A6M0001485, A6M0001486, A6M0001498

Distribution pattern: Us and Canada

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    During internal cycle testing of A-Series lifts Handicare has found that the A-Series lift strap for lifts manufactured between April 25, 2017 and June 12, 2017 may wear prematurely at max load (625 lbs). We have identified the root cause as a supplied component that was not to specification. If the strap wears during use there is a potential risk for the person being supported by the lift to drop.