Brilliance CT Big Bore Oncology, Model number 728243 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes.
Z-2653-2016
Product summary
- Event
- Event 74820
- Status
- Terminated
- Classification
- Class II
- Quantity
- 33
- Official record key
device-enforcement:Z-2653-2016
Official wording
Reason: After upgrading to 3.6.7 software version via FCO72800643, during reconstruction of gated helical scans, the planned anatomy can be cut off on the CT images. This issue can occur on gated CT helical reconstruction that is not planned at iso-center (0,0). If the region of interest is not visualized in the images, a CT rescan may be performed.
Code information: 728243 Brilliance BigBore Oncology CT 7116 7611 7354 7063 7309 7169 7205 7016 7019 7043 7069 7073 7091 7112 7132 7163 7141 7194 7226 7325 7294 7350 7316 7480 7114 7210 7310 7438 7530 7186 7119 7042
Distribution pattern: US Nationwide Distribution in the states of: AL,AR,AZ,CA,CO,CT,DC,DE,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA, MD,ME,MI,MN,MO,MS,NC,ND,NE,NJ,NM, NV,NY,OH,OK,OR,PA,PR,SC,TN,TX,UT,VA,VT,WA,WI,WV,WY
Derived failure modes
-
Device software or design
software version via FCO72800643, during reconstruction of gated helical scans, the planned anatomy can