Skip to content
Recall Observatory FDA recall evidence

Device product

MCA7-1SD - 1MM Diamond Ball, 7.3CM. All lots are included in recall. Intended for use in cutting and shaping delicate bone material, primarily in otologic procedures such as cochleostomy.

Z-0767-2014

March 29, 2012

Class II

Product summary

Firm
The Anspach Effort, Inc.
Event
Event 63624
Status
Terminated
Classification
Class II
Quantity
6439 total
Official record key
device-enforcement:Z-0767-2014

Official wording

Reason: The Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be broader than the validation of the MCA Cutting Burrs and its directions for use did not clearly state the correct intended use of the device. After a labeling correction, Anspach Effort, Inc. decide

Code information: All lots are included in recall

Distribution pattern: Worldwide Distribution: US (nationwide) and Internationally to: AU, BE, CA, CH, ES, FR, GB, IT, JP, NL, SE, and ZA.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be broader than the validation of the MCA Cutting Burrs and its directions for use did not clearly state the correct intended use of the device. After a labeling correction, Anspach Effort, Inc. decide