Skip to content
Recall Observatory FDA recall evidence

Device product

System 1 Base 100/120V The Terumo Advanced Perfusion System 1 is indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment.

Z-1970-2012

June 25, 2012

Class II

Product summary

Firm
Terumo Cardiovascular Systems Corporation
Event
Event 62373
Status
Terminated
Classification
Class II
Quantity
992 units
Official record key
device-enforcement:Z-1970-2012

Official wording

Reason: Terumo Cardiovascular Systems (Terumo CVS) has received reports of various roller pump malfunctions of the Terumo Advanced Perfusion System 1 that could result in the inability to use the pump and would then require replacement with back-up equipment. Description of malfunctions: " Unplanned pump stop or pause " Changes in pump speed If it is necessary to replace a roller pump during cardiopu

Code information: catalog number: 801763 and serial numbers: 0011-0335, 0337-0610, 1001-1034, 1037-1049, 1100, 1103-1106,1109-1445, and 1447-1450.

Distribution pattern: Worldwide Distribution-USA (nationwide) and the countries of ARGENTINA, AUSTRALIA, BELGIUM, CANADA, CHILE, COLOMBIA, COLOMBIA, Costa Rica, Dominican Republic, Egypt, Germany, Guatamala, Honduras, Hong Kong, India, Indonesia, Israel, Japan, Jordan, Kuwait, Malaysia, Mexico, Pakistan, Philippines, Russia, Saudi Arabia, Singapore, South Africa, South Korea, Taiwan, Thailand, Turkey, UNITED ARAB EMIRATES (UAE), and Vietnam.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Terumo Cardiovascular Systems (Terumo CVS) has received reports of various roller pump malfunctions of the Terumo Advanced Perfusion System 1 that could result in the inability to use the pump and would then require replacement with back-up equipment. Description of malfunctions: " Unplanned pump stop or pause " Changes in pump speed If it is necessary to replace a roller pump during cardiopu