Skip to content
Recall Observatory FDA recall evidence

Device product

BIOMET Spine, 5.5 TI 5.5x50mm Translation Screw Polaris 5.5 TI Spinal System. Rx Only. Non-Sterile. Intended for immobilization and stabilization as an adjunct to fusion as a pedicle screw fixation system

Z-0359-2014

August 22, 2013

Class II

Product summary

Firm
Ebi, Llc
Event
Event 66300
Status
Terminated
Classification
Class II
Quantity
7,092 units
Official record key
device-enforcement:Z-0359-2014

Official wording

Reason: Polaris Translation Screw Failures were recently reported to Biomet Spine. These failures are generally described as dissociation of the screw head from the shaft of the Polaris Translation Screw.

Code information: Multiple Item and Lot numbers. The firm expects all Translation Screw Systems (containing a cobalt chrome component) used and contacted with the SpF Implantable Spine Fusion Stimulator (DC electrical current) to be at risk for this failure mode.

Distribution pattern: Nationwide distribution including Puerto Rico.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0