Device product

BRAVO PH CAPS DELIVERY DEV5-PK, FGS-0312/FGS-0313 - Product Usage: intended to be used for gastroesophageal pH measurement and monitoring of gastric reflux in adults and children from 4 years of age.

Z-1335-2021

February 26, 2021

Class II

Product summary

Firm
Covidien Llc
Event
Event 87396
Status
Ongoing
Classification
Class II
Quantity
62551
Official record key
device-enforcement:Z-1335-2021

Official wording

Reason: The capsule may fail to attach to the esophageal mucosa, which in turn may lead to the aspiration of the capsule. In the event of capsule aspiration, immediate intervention to remove the capsule is required. Potential outcomes following capsule aspiration include low oxygen saturation, intervention to retrieve the capsule, potential need to intubate the patient, extended hospital stay and a delay of treatment. Thirteen reports of capsule aspiration have been received the last 2 years.

Code information: 49349Q 49350Q 49351Q 49352Q 49799Q 49800Q 49801Q 49802Q 49809Q 49849Q 49850Q 49851Q 50392Q 50393Q 50492Q 50493Q 50494Q 50495Q 50496Q 50497Q 50615Q 50616Q 50868Q 50869Q 50870Q 50871Q 50872Q 50873Q 50874Q 50875Q 50876Q 50877Q 50878Q 50879Q 50880Q 50881Q

Distribution pattern: Worldwide distribution - US Nationwide distribution.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The capsule may fail to attach to the esophageal mucosa, which in turn may lead to the aspiration of the capsule. In the event of capsule aspiration, immediate intervention to remove the capsule is required. Potential outcomes following capsule aspiration include low oxygen saturation, intervention to retrieve the capsule, potential need to intubate the patient, extended hospital stay and a delay of treatment. Thirteen reports of capsule aspiration have been received the last 2 years.