Device product
VS II Aliquoter System calculator/data processing module, for clinical use Product Usage: Elecysy 411 analyzer is a fully automated, random access, computer controlled analytical systems for quantitative and qualitative determinations of analytes in body fluids.
Z-0103-2017
Product summary
- Event
- Event 75186
- Status
- Terminated
- Classification
- Class II
- Quantity
- 9014 in total
- Official record key
device-enforcement:Z-0103-2017
Official wording
Reason: Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers sample rack tray. The cleaning procedure for the sample rack tray has been revised to clarify the cleaning process.
Code information: Not applicable
Distribution pattern: US Nationwide Distribution
Derived failure modes
-
Unknown
Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers sample rack tray. The cleaning procedure for the sample rack tray has been revised to clarify the cleaning process.