Skip to content
Recall Observatory FDA recall evidence

Device product

PSG-1100 Sleep Diagnostic System. Sample product label: JB-110A SERIAL No. 01454 RoHS Compliant; PSG-100 Head Box; UDI (01) 04931921108383; Made in Japan The PSG-1100 Sleep Diagnostic System is a digital PSG amplifier intended to record the electrical activity of the brain (EEG) and other bio-potential signals and to record physiological data required for sleep studies.

Z-0865-2021

December 21, 2020

Class II

Product summary

Firm
Nihon Kohden America Inc
Event
Event 87011
Status
Ongoing
Classification
Class II
Quantity
189 Units
Official record key
device-enforcement:Z-0865-2021

Official wording

Reason: Medical device non-conformance to electrical safety standard (IEC 60601-1).

Code information: Serial numbers 3. Serial numbers 320, 470, 507, 534, 621, 757, 770, 795, 892, 1165-1183, 1186- 1205, 1214-1293, 1295-1297, 1304-1330, 1333-1342, 1358 -1384, 1387-1389, 1391-1413, 1421-1430, 1434-1505, 1507- 1516.

Distribution pattern: US Nationwide Distribution - CA, CO, CT, FL, ID, IL, IN, MA, MD, MI, MO, NC, NM, NY, SC, TX, TN, UT, VA, VT and WA.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Medical device non-conformance to electrical safety standard (IEC 60601-1).