Skip to content
Recall Observatory FDA recall evidence

Device product

Baxter SIGMA Spectrum Infusion Pump with Master Drug Library, Model #/Product Code 35700BAX2

Z-1693-2017

January 26, 2017

Class II

Product summary

Firm
Baxter Healthcare Corporation
Event
Event 76364
Status
Terminated
Classification
Class II
Quantity
Domestic: 14,175 units; Canada: 187 units
Official record key
device-enforcement:Z-1693-2017

Official wording

Reason: SIGMA Spectrum Infusion Pumps with Master Drug Library, Software V8.00.02 contains an anomaly that may cause a Sharp Watchdog Timeout error message. Specifically, the software anomaly causes the error message only under conditions in which a message is being written to the Spectrum Infusion Pumps Event History Log (EHL), and the length of that message exactly matches the size of the available memory at the end of a memory sector. For this reason, the issue is significantly more likely to manifest when poor network maintenance results in high volumes of messages being written to the EHL.

Code information: 14,362 Serial Numbers (please contact CDRH for list of affected Serial Numbers).

Distribution pattern: Natiowide, Puerto Rico and Canada

Derived failure modes

  • Device software or design

    reason.device_software_design · v1.0.0

    Software V8.00.02 contains an anomaly