AORTIC ARCH CANNULA, Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass (part numbers NA-55X7 and NA-55X8) Aortic Arch Cannulae are designed to be used in the extracorporeal circuit during cardiopulmonary bypass surgery. The cannula consists of varying lengths of non-wire reinforced (flexible) polyvinyl chloride tubing that terminates in an angled tip. These devices come in direct contact with the central circulatory system but they are not intended to control, diagnose, monitor or correct a defect.
Z-2147-2016
Product summary
- Event
- Event 74490
- Status
- Terminated
- Classification
- Class II
- Quantity
- 5740
- Official record key
device-enforcement:Z-2147-2016
Official wording
Reason: Sorin Group USA is recalling Aortic Arch Cannulae (part numbers NA-55X7 and NA-55X8) because the distal end may be less than specified. The defective cannula are showing an angle that is about 15-20 degrees instead of approximately 45 degrees.
Code information: Lots: 1505600017 ; 1505600018 ; 1506500005 ; 1509200078 ; 1513100112 ; 1513900027 ; 1513900028 ; 1514700045 ; 1516600192 ; 1516800126 ; 1517800004 ; 1519500037 ; 1520100049 ; 1520100050 ; 1520900031 ; 1520900033 ; 1521600126 ; 1523000093 ; 1523600120 ; 1525100023 ; 1525800026 ; 1526500032 ; 1526500033 ; 1526500034 ; 1526500072 ; 1527200050 ; 1527200051 ; 1527200052 ; 1527800163 ; 1529300026 ; 1529300027 ; 1529300028 ; 1533500033 ; 1534200034 ; 1601900027 ; 1601900028 ;
Distribution pattern: Worldwide Distribution: US (nationwide) to states of: LA, Ml, MO, OH, TN, and TX; and countries of: Italy and Canada.
Derived failure modes
-
Unknown
Sorin Group USA is recalling Aortic Arch Cannulae (part numbers NA-55X7 and NA-55X8) because the distal end may be less than specified. The defective cannula are showing an angle that is about 15-20 degrees instead of approximately 45 degrees.