Skip to content
Recall Observatory FDA recall evidence

Device product

Product 9 consists of all product under HRS, and same usage: Item no: 47494502601 4.5 BROAD SCP PLT, 26 HOL Product Usage: Temporary internal fixation devices designed to stabilize fractures during the normal healing process.

Z-1107-2016

January 11, 2016

Class II

Product summary

Firm
Zimmer Manufacturing B.V.
Event
Event 72900
Status
Terminated
Classification
Class II
Quantity
0
Official record key
device-enforcement:Z-1107-2016

Official wording

Reason: As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.

Code information: lot no: 61889853

Distribution pattern: US Nationwide Distribution

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.