Device product

MEE-1000A Neuromaster, Model No. Accessories: MS-110B (Electric Stimulation), MS-120BK (Current Stimulator Box) JB-116B, JB-132B (Amplifiers) Electromyography. (EMG)

Z-0817-2016

January 19, 2012

Class II

Product summary

Firm
Nihon Kohden America Inc
Event
Event 73114
Status
Terminated
Classification
Class II
Quantity
477 units total (137 units in US)
Official record key
device-enforcement:Z-0817-2016

Official wording

Reason: The MEE-1000A Neuromaster had defective printed circuit boards (PCBs). Gas arrestors on these boards which were intended to provide protection from electrostatic discharge failed to prohibit unintentional current flow.

Code information: Serial Numbers: MS-110B: (17 units) 00013-00019, 00023- 00032 MS-120BK: (6 units) 0009- 00014 JB-116B: (31 units), 0008, 00012, 00014, 00016, 00018, 00019, 00020, 00021, 00023, 00025, 00026, 00027, 00028, 00029, 00031, 00032, 00034, 00035, 00039, 00043, 00102,00103, 00122-00128, 00134, 00135 JB-132B: (83 units) 00009, 00011, 00012, 00014, 00015, 00016, 00018, 00019, 00020, 00021, 00023, 00024, 00025, 00026, 00027, 00036, 00037, 00038, 00041, 00042, 00043, 00044, 00046, 00047, 00049, 00061, 00062, 00066, 00067, 00068, 00069,00071, 00072, 00074, 00075, 00082, 00084, 0085, 00086, 00090, 00095, 00099, 00120, 00121-00129, 00131-00147, 00149-00155, 00157, 00161, 00164-00168

Distribution pattern: Worldwide distribution. US nationwide, Australia, Austria, France, Germany, Greece, Japan, Romania, South Africa, South Korea, Spain, and United Kingdom.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The MEE-1000A Neuromaster had defective printed circuit boards (PCBs). Gas arrestors on these boards which were intended to provide protection from electrostatic discharge failed to prohibit unintentional current flow.