Device product

Hitachi Scenaria CT System Software Product Usage: The Scenaria system is indicated to acquire axial volumes of the whole body including the head. The images can be acquired in either axial, helical, gated or dynamic modes. The volume datasets acquired by the Scenaria can be post processed by the Scenaria to provide additional information. Post processing capabilities included in the Scenaria software include CT angiography (CTA), Multi-planar reconstruction (MPR) and volume rendering.

Z-2037-2012

May 04, 2012

Class II

Product summary

Firm
Hitachi Medical Systems America Inc
Event
Event 62372
Status
Terminated
Classification
Class II
Quantity
Six (6) units
Official record key
device-enforcement:Z-2037-2012

Official wording

Reason: Hitachi discovered a software error in the Reconstruction Status function that creates a potential risk that a patient study will not be fully reconstructed. When multiple reconstructions are queued for processing, if a queue is suspended and the task order changed, the second queue may not reconstruct the full range of slices. This may result in missing diagnostic data.

Code information: Serial #'s S5001 - S5011

Distribution pattern: US Nationwide Distribution to the following states of: MD, MI, MX, ND, NY, OH and TX.

Derived failure modes

  • Device software or design

    reason.device_software_design · v1.0.0

    software error