Skip to content
Recall Observatory FDA recall evidence

Device product

SENSATION PLUS¿ 8Fr. 50cc Intra-Aortic Balloon Catheter with Insertion Kit & STATLOCK¿ Product Code/Part Number: 0684-00-0576-01U is a cardiac assist device which support the heart's left ventricle by increasing coronary perfusion and reducing the work of the left ventricle.

Z-2164-2017

April 12, 2017

Class II

Product summary

Firm
Datascope Corporation
Event
Event 76942
Status
Terminated
Classification
Class II
Quantity
996 UNITS
Official record key
device-enforcement:Z-2164-2017

Official wording

Reason: Maquet/Getinge is initiating a voluntary product removal involving four lot numbers of SENSATION PLUS 8Fr. 50cc Intra-Aortic Balloon Catheter with Insertion Kit & STATLOCK since during internal pressure (leak) testing of the SENSATION PLUS 8Fr. 50cc, IABC, failures related to the tip seal leak were identified.

Code information: 300004515, 3000043376 3000040272 3000038372 added lot numbers: 3000043374 and 3000045513

Distribution pattern: Nationwide

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Maquet/Getinge is initiating a voluntary product removal involving four lot numbers of SENSATION PLUS 8Fr. 50cc Intra-Aortic Balloon Catheter with Insertion Kit & STATLOCK since during internal pressure (leak) testing of the SENSATION PLUS 8Fr. 50cc, IABC, failures related to the tip seal leak were identified.