Device product
CyberKnife VSI, Part Numbers: a) 032500-010: CyberKnife VSI 9.6.x Robotic Radiosurgery System for Japan b) 033000-010: CyberKnife VSI Robotic Radiosurgery System
Z-1697-2019
Product summary
- Event
- Event 82664
- Status
- Terminated
- Classification
- Class II
- Quantity
- 43 units
- Official record key
device-enforcement:Z-1697-2019
Official wording
Reason: There is an unintended drop of a secondary collimator housing that can occur during housing Xchange(TM) workflows. Specifically, during an automated housing Xchange, an interlock may interrupt this operation causing the collimator housing to partially attach.
Code information: a) 032500-010: CyberKnife VSI 9.6.x Robotic Radiosurgery System for Japan Serial Numbers: C0281, C0307 b) 033000-010: CyberKnife VSI Robotic Radiosurgery System Serial Numbers: C0042, C0064, C0071, C0074, C0079, C0091, C0098, C0102, C0137, C0161, C0180, C0182, C0187, C0200, C0203, C0210, C0216, C0230, C0238, C0246, C0247, C0249, C0252, C0255, C0257, C0264, C0269, C0270, C0273, C0274, C0276, C0277, C0284, C0293, C0294, C0304, C0305, C0322, C0323, C0325, C0326
Distribution pattern: US & Italy, United Kingdom, Switzerland, Belgium, Canada, Germany, France, Japan, Poland, Netherlands, Luxembourg, Australia, Qatar, Latvia, Portugal, India, Hungary, Turkey, United Arab Emirates, Thailand
Derived failure modes
-
Unknown