Skip to content
Recall Observatory FDA recall evidence

Device product

Philips Ultrasound, Model Q-Station, with software version 3 or higher, Catalog number: 795088; Part Number: 989605382391 Q-Station is application software intended to manage, view, analyze, and report qualitative and quantitative image data from ultrasound exams. It is designed to host optional advanced analysis applications via QLAB integration and provide integrated tools that allow users to manually assess and score cardiac wall motion and export images and / or exams and reports. Q-Station can view DICOM images of non-ultrasound images such as CT, MR, NM, CR, MG, XA, PET, RT and X-Ray modalities for reference viewing. It support connectivity to ultrasound systems, PACS and other DICOM storage repositories.

Z-1807-2015

May 15, 2015

Class II

Product summary

Firm
Philips Ultrasound, Inc.
Event
Event 71432
Status
Terminated
Classification
Class II
Quantity
125 units (30 in US and 95 international)
Official record key
device-enforcement:Z-1807-2015

Official wording

Reason: Philips Healthcare has discovered a problem in the Philips Ultrasound Q-Station version 3.0 or higher that could result in measurements from a study (Structured Report) to be appended to subsequent studies for other patients.

Code information: *****************US Serial numbers***************************************** 2176-IAF08E60-15D4, 2187-VAC5T6CD-157D, 2C98-25A23E3A-4B2E, 34F5-4084F543-80EA, 3A5D-4079F60B-80D4, 54B0-HADF24F2-15B1, 678B-NAE42F41-15BB, 6FF9-259B5EB7-4B20, 725B-GACE30D1-158F, 72EE-408914F8-80F2, 7A39-497627F7-92C9, 8536-IAD33B20-1599, A584-QAE84EF6-15C3, A838-IAE2CF5A-15B8, B903-407CB5D9-80D9, BDC7-TAE25A0D-15B7, D356-YAE1E4C0-15B6, D60A-GADC6524-15AB, D8BE-KAD6E588-15A0, DB72-RAD165EC-1595, DE26-WACBE650-158A, E0DA-XAC666B4-157F, E16D-40814ADB-80E2, E8E5-TAE16F73-15B5, EE4D-IAD6703B-159F, F669-PAC5F167-157E, F90C-MAEBF95E-15CA, H944-VACAFBB6-1589, MBF8-WAC57C1A-157E, U9D7-4085DFDD-80EC. ********************Serial numbers in CANADA**************************************** 1ED3-QACA8669-1588, 4489-SAD49B07-159C, B054-UAD25086-1597, C897-TACC5B9D-158B, and F3B5-IACB7103-1589. *********************Serial numbers in other countries*********************************** 114F-QAE6M475-15BF, 11F3-40756109-80CB, 16B7-KADBT53D-15A9, 196B-UAD585A1-159E, 1F66-40856A90-80EB, 221A-407FEAF4-80E0, 23FD-4977FD2B-92CD, 242A-MAEBPEC4-15C9, 24CE-407A6B58-80D5, 2782-4074EBBC-80CA, 2965-496CFDF3-92B7, 29E4-25A7BDD6-4B39, 2C46-QADA8FF0-15A8, 2EFA-RAD51054-159D, 2F4C-259CBE9E-4B23, 3716-IAC49180-157C, 37A9-407F75A7-80DF, 3C6D-XAE519DB-15BD, 3CBF-25ACC825-4B43, 3D11-4074766F-80C9, 41D5-NADA1AA3-15A7, 47D0-4089FF92-80F4, 49F1-IAC99BCF-1586, 4C94-YAEFA3C6-15D2, 4CA5-UAC41C33-157B, 4F1B-49771291-92CB, 4FEC-407980BE-80D3, 524E-25AC52D8-4B42, 5764-MAD9A556-15A6, 5A18-QAD425BA-159B, 5CCC-HACEA61E-1590, 5F80-MAC92682-1585, 6013-4084IAA9-80E8, 61F6-497C1CE0-92D5, 62C7-407E8B0D-80DD, 657B-4079TB71-80D2, 6CF3-WAD93009-15A5, 750F-TAC8B135-1584, 75A2-4083955C-80E7, 7856-407E15C0-80DC, 7B0A-40789624-80D1, 7CED-4970A85B-92BE, 7D6C-25AB683E-4B40, 7FCE-HADE3A58-15AF, 8282-JAD8BABC-15A4, 87EA-PACDBB84-158E, 8A9E-RAC83BE8-1583, 8B31-4083200F-80E6, 8D14-497B3246-92D3, 8D41-NAEE43DF-15CF, 8DE5-407DA073-80DB, 8FF5-NAE8C443-15C4, 9099-407820D7-80D0, 9811-WAD8456F-15A3, 9D79-WACD4637-158D, A0C0-4082AAC2-80E5, A374-407D2B26-80DA, A628-4077AB8A-80CF, A80B-496FBDC1-92BC, AAEC-MADD4FBE-15AD, ADA0-PAD7D022-15A2, B308-JACCD0EA-158C, B39B-4087B511-80EF, B5BC-RAC7514E-1581, B64F-40823575-80E4, B832-497A47AC-92D1, B85F-UAED5945-15CD, BAE6-4974C810-92C6, BBB7-4077363D-80CE, BE19-25AAH857-4B3D, C07B-KADCDA71-15AC, CBDE-4081C028-80E3, CDC1-4979D25F-92D0, CE92-407C408C-80D8, D146-4076C0F0-80CD, E37D-JAEC6EAB-15CB, E421-407BCB3F-80D7, F448-4086552A-80EC, F6FC-4080D58E-80E1, F8DF-4978E7C5-92CE, F9B0-407B55F2-80D6, FC64-4075D656-80CB, T3DC-SAD5FAEE-159F, TF3F-407AE0A5-80D6, VC8B-40806041-80E1, and Y690-GAD07B52-1594. **********NEWLY ADDED FOUR AFFECTED SERIAL NUMBERS******* aaec-madd4fbe-15ad; WE9B-QAEB8411-15CA; 2B14-MDE9A2F9-1BC6; and 3C40-4972P842-92C1

Distribution pattern: Worldwide Distribution - US (nationwide) and Puerto Rico; internationally to Australia, Canada, Colombia, France, Germany, Great Britain, India, Italy, Mexico, Netherlands, Norway, Pakistan, Poland, Russia, Singapore, Spain, Switzerland, Thailand, and Zimbabwe.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Philips Healthcare has discovered a problem in the Philips Ultrasound Q-Station version 3.0 or higher that could result in measurements from a study (Structured Report) to be appended to subsequent studies for other patients.