Device product
Triage D-dimer PN 98100 Used as an aid in the assessment and evaluation of patients suspected of having disseminated intravascular coagulation or thromboembolic events including pulmonary embolism.
Z-1932-2012
Product summary
- Event
- Event 61927
- Status
- Terminated
- Classification
- Class I
- Quantity
- 1,972 kits
- Official record key
device-enforcement:Z-1932-2012
Official wording
Reason: Certain lots of the affected products may have significantly decreased precision relative to the package insert, which could result in an increased frequency of false positive or false negative results. These false positive and false negative results are unpredictable within lots and may not be detected by Quality Control testing.
Code information: Phase 1: W50069B, W50070B, W50071B, W50541B, W50551B, W50552B, and W50556B. Phase 2: W50550B
Distribution pattern: Nationwide Distribution
Derived failure modes
-
Unknown
Certain lots of the affected products may have significantly decreased precision relative to the package insert, which could result in an increased frequency of false positive or false negative results. These false positive and false negative results are unpredictable within lots and may not be detected by Quality Control testing.