Medtronic, Viva Quadripolar Implantable Cardioverter Defibrillators: Product Model VIVAXTCRT-D DTBA1D1, DTBA1D4 VIVASCRT-D DTBB1D1, DTBB1D4 Product Usage: The Medtronic Viva CRT -D dual chamber, implantable cardioverter defibrillator with cardiac resynchronization therapy (CRT -D) is a multiprogrammable cardiac device that monitors and regulates the patients heart rate by providing single or dual chamber, rate responsive bradycardia pacing; sequential biventricular pacing; ventricular tachyarrhythmia therapies; and atrial tachyarrhythmia therapies.

Z-2650-2016

August 12, 2016

Class II

Product summary

Firm
Medtronic Inc., Cardiac Rhythm and Heart Failure
Event
Event 74923
Status
Terminated
Classification
Class II
Quantity
78 (39 US, 39 OUS)
Official record key
device-enforcement:Z-2650-2016

Official wording

Reason: 78 Viva CRT -Ds and Evera ICDs that were manufactured with a specific subset of circuit components may experience rapid battery depletion due to a low resistance path developing within the circuit component. There have been seven (7) confirmed failures (9%) through July 16, 2016. Reported complications have included shortness of breath, pocket heating, low heart rate, and early device explant.

Code information: Product Name: Viva XT CRT -D . Models: DTBA1D4, DTBA1D1. Serial Numbers: BLF203128H, BLF204746H, BLE202888H, BLE202889H, BLE202890H, BLE202901H, BLE202941H, BLE202947H, BLE202954H, BLE202958H, BLE202961H, BLE202962H, BLE202964H, BLE202981H, BLE202987H, BLE202989H, BLE202990H, BLE202991H, BLE203019H, BLE203026H, BLE203027H, BLE203029H, BLE203032H, BLE203046H, BLE203052H, BLE203073H Product Name: Viva S CRT-D. Models: DTBB1D4, DTBB1D1. Serial Numbers: BLO202272H, BLN202206H.

Distribution pattern: Worldwide Distribution - US Nationwide in the states of US: AZ, CA, FL, GA, IL, KY, MD, MN, NJ, NY, NC, OH, OK, PA, PR, TX. and countries of: Australia, Finland, France, Germany, Iran, Italy, Japan, Korea, Netherlands, Puerto Rico, Saudi Arabia, Spain, and United Kingdom.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    78 Viva CRT -Ds and Evera ICDs that were manufactured with a specific subset of circuit components may experience rapid battery depletion due to a low resistance path developing within the circuit component. There have been seven (7) confirmed failures (9%) through July 16, 2016. Reported complications have included shortness of breath, pocket heating, low heart rate, and early device explant.