Skip to content
Recall Observatory FDA recall evidence

Device product

NICO Myriad Console Model number NN-7000 The Myriad System is a surgical system specifically designed for morcellation of soft tissue. It primarily consists of a reusable console with foot pedal and a disposable handpiece. For the morcellation and removal of tissue during pelviscopic, laparoscopic, percutaneous and open surgical procedures whenever access to the surgical site is limited.

Z-2767-2015

July 29, 2015

Class II

Product summary

Firm
Nico Corp.
Event
Event 72015
Status
Terminated
Classification
Class II
Quantity
6 units
Official record key
device-enforcement:Z-2767-2015

Official wording

Reason: Two cables within the console were inadvertently twisted during manufacturing. The individual wires within the cables may break and/or contact one another where the cables terminate into two connectors. When these wires break or contact one another (i.e., short), various types of malfunction may occur with the console. One type of malfunction may result in unexpected aspiration increase.

Code information: Model number NN-7000 SN: 00231, 00232, 00242, 00245, 00248, 00250

Distribution pattern: US Distribution to states of: MD, IN, CA, and NC.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Two cables within the console were inadvertently twisted during manufacturing. The individual wires within the cables may break and/or contact one another where the cables terminate into two connectors. When these wires break or contact one another (i.e., short), various types of malfunction may occur with the console. One type of malfunction may result in unexpected aspiration increase.