Skip to content
Recall Observatory FDA recall evidence

Device product

Lactate Membrane units for E7077 Electrode Radiometer Medical ApS. The intended use is in vitro testing of samples of whole blood.

Z-2303-2013

October 08, 2012

Class II

Product summary

Firm
Radiometer America Inc
Event
Event 66108
Status
Terminated
Classification
Class II
Quantity
13,632 boxes/4 membranes
Official record key
device-enforcement:Z-2303-2013

Official wording

Reason: RADIOMETER became aware that some membranes may have enzyme residue on the outer membrane. The enzyme residue may cause an initial negative bias on the reported Lactate result upon replacement of the Lactate membrane.

Code information: Part Number: 942-066 R1741 to R1869

Distribution pattern: Nationwide Distribution, including the states of: AK, AL, AR, AZ, BM, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI and WV.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0