EasyLink(TM) Informatics System, Part Number 1000034805. EasyLink(TM) Informatics System is classified as a data calculator and processing module for use with In-Vitro diagnostics devices.

Z-2172-2012

June 07, 2012

Class II

Product summary

Firm
Siemens Healthcare Diagnostics, Inc.
Event
Event 62448
Status
Terminated
Classification
Class II
Quantity
237
Official record key
device-enforcement:Z-2172-2012

Official wording

Reason: There is a potential for the EasyLink system to display and transmit to the Laboratory Information System networks an incorrect result following the EasyLink systems application of the configured result number format. The problem is encountered when the result number format configured on the Sysmex CA-7000 or CA-1500 coagulation analyzer does not match the EasyLink system settings on the following

Code information: Version 5.0 and Version 5.0 Service Pack 1 through 4

Distribution pattern: Worldwide distribution, including the countries of Australia, Austria, Belgium, Canada, Czech Republic, France, Germany, Italy, Netherlands, New Zealand, Slovakia, Spain, Switzerland and the United States, with nationwide distribution. Center Recall Depth is Retail.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    There is a potential for the EasyLink system to display and transmit to the Laboratory Information System networks an incorrect result following the EasyLink systems application of the configured result number format. The problem is encountered when the result number format configured on the Sysmex CA-7000 or CA-1500 coagulation analyzer does not match the EasyLink system settings on the following