SybronEndo Replacement AC Power Cord, US Elements Obturation; REF 973-0334; SybronEndo, 1332 S. Lone Hill Ave. Glendora, CA 91740; EU Rep: Kerr Italia, S.r.l, Via Passanti, 332, 1-84018 Scafati (SA), Italy, Made in Taiwan The intended use of the Elements Obturation Unit and Replacement A/C Power Cord is to provide continuous heat at the tip of the dental instrument to test tooth response to thermal stimulus, for tissue cauterization and coagulation and to backfill and down pack gutta percha during endodontic root canal treatment.

Z-1545-2013

January 17, 2013

Class II

Product summary

Firm
Ormco/Sybronendo
Event
Event 64216
Status
Terminated
Classification
Class II
Quantity
3615 unit total
Official record key
device-enforcement:Z-1545-2013

Official wording

Reason: A recall has been issued by SybronEndo because a Canadian Standards Association (CSA)-directed cautionary statement label was inadvertently omitted from the Elements Obturation Unit and its Replacement A/C Power Cord.

Code information: Lot/Serial Numbers: 3757349, 4220920, 4304917, 4367699, 436513, 4445719, 4458262, 4522605, 4581118, 4724407, 4730212

Distribution pattern: Worldwide Distribution - USA (nationwide) and the countries of Australia, Azerbaijan, Belgium, Canada, China, Cyprus, Czech Republic, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, Iceland, India, Italy, Japan, Kuwait, Latvia, Lebanon, Malaysia, Netherlands, New Zealand, Pakistan, Poland, Romania, Russia, Saudi Arabia, Singapore, Spain, Sweden, Switzerland, Thailand, Turkey, Ukraine, United Arab Emirates, and United Kingdom.

Derived failure modes

  • Labeling or packaging

    reason.labeling_packaging · v1.0.0

    label was inadvertently omitted