Skip to content
Recall Observatory FDA recall evidence

Device product

CROSSCHECK(R) Plating system, MTP NX Plate, Right, Recon, REF CCP-MPN2R. Orthopedic use.

Z-1491-2015

March 06, 2015

Class II

Product summary

Firm
Wright Medical Technology, Inc.
Event
Event 70859
Status
Terminated
Classification
Class II
Quantity
301 units
Official record key
device-enforcement:Z-1491-2015

Official wording

Reason: The minor diameter, major diameter, thread pitch, and pitch on the Crosscheck are out of specification, which may result in intraoperative malfunction

Code information: Lot Numbers: 3504919, 35049-19A, 35049-19A-1, 35049-20, 362792

Distribution pattern: Nationwide Distribution.

Derived failure modes

  • Potency or specification failure

    reason.potency_specification_failure · v1.0.0

    out of specification