Device product
Puritan Bennett 980 Ventilator System, Universal (with neonatal functionality enabled) and Neonatal Models. Designed for use on Neonatal (NICU) through Adult patient populations who require respiratory support or mechanical ventilation and weigh a minimum of 0.3 kg (0.66 lb).
Z-2329-2015
Product summary
- Event
- Event 71675
- Status
- Terminated
- Classification
- Class I
- Quantity
- 657 units
- Official record key
device-enforcement:Z-2329-2015
Official wording
Reason: Reports in which tidal volumes reaching patients were lower than set tidal volumes in neonatal Volume Control Plus (VC+) Mode with active humidification. This situation may potentially lead to respiratory compromise if not recognized.
Code information: PB980 Ventilator (980xxxxxxxx): 980U1ENDIUU, 980U1ENDIUUS, 980U3ENDIUU, 980U3ENDIUUS, 980N1ENDIUU, 980N1ENDIUUS, 980N3ENDIUU, 980N3ENDIUUS
Distribution pattern: Worldwide Distribution-US (nationwide) and the countries of Canada, United Arab Emirates, Argentina, Australia, Bahrain, Brazil, Chile, Colombia, Costa Rica, Egypt, Ireland, Israel, Jordan, Japan, Kuwait, Lebanon, Morocco, Mexico, Panama, Peru, Qatar, Russia, Saudi Arabia, Singapore, Slovenia, South Africa, Turkey, and Uruguay.
Derived failure modes
-
Unknown
Reports in which tidal volumes reaching patients were lower than set tidal volumes in neonatal Volume Control Plus (VC+) Mode with active humidification. This situation may potentially lead to respiratory compromise if not recognized.