Device product

Fluorescence Imaging Procedure Kit, designed for use with Firefly Fluorescence Imaging for the da Vinci Xi 4000 Surgical Systems. The Intuitive Surgical Endoscopic Instrument Control System (Intuitive Surgical da Vinci Surgical System Model IS3000) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments.

Z-0835-2015

November 21, 2014

Class II

Product summary

Firm
Intuitive Surgical, Inc.
Event
Event 69845
Status
Terminated
Classification
Class II
Quantity
1532
Official record key
device-enforcement:Z-0835-2015

Official wording

Reason: The storage conditions for indocyanine green (ICG) are not on the individual Fluorescence Imaging Procedure Kit box, the six-kit box, and the shipping box. If heat were to damage the material, the result would be a loss of fluorescence.

Code information: PN 950156-02

Distribution pattern: Nationwide Distribution.