Device product
Continuous Ventilator/Continuous Ventilator with Respiratory Mechanics, Model No. V1000
Z-2084-2013
Product summary
- Event
- Event 48997
- Status
- Terminated
- Classification
- Class II
- Quantity
- 3,255 units (2302 units in US)
- Official record key
device-enforcement:Z-2084-2013
Official wording
Reason: The power to the Esprit Display's Backlight is interrupted, causing the GUI to flicker or go dark and making the display unreadable.
Code information: Model Number: V1000
Distribution pattern: USA Nationwide and Worldwide to Algeria, Argentina, Australia, Bangladesh, Brazil, Cambodia, Canada, Chile, Colombia, Ecuador, Egypt, France, Germany, Guatemala, Hong Kong, India, Italy, Japan, Libya, Malaysia, Mexico, Pakistan, Peru, Philippines, Puerto Rico, Russia, South Africa, South Korea, Spain, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Venezuela, and Vietnam.
Derived failure modes
-
Unknown
The power to the Esprit Display's Backlight is interrupted, causing the GUI to flicker or go dark and making the display unreadable.