Device product
MicroScan WalkAway-40 plus Instrument and MicroScan Walkaway-96 plus instrument access door hinge. Siemens Healthcare Diagnostics, Inc. in vitro diagnostic
Z-1990-2014
Product summary
- Event
- Event 68524
- Status
- Terminated
- Classification
- Class II
- Quantity
- WA40: 560 US, 131 OUS, 9 already repaired; WA96: 81 US, 189 OUS, 57 repaired.
- Official record key
device-enforcement:Z-1990-2014
Official wording
Reason: Springs contained in the access door hinge assembly on the Walk Away Plus instruments are becoming detached causing premature failure of the hinge assembly and a potential safety issue for the customers.
Code information: WalkAway 40 - Material number: 10444853; catalog number: B1018-283; WalkAway 96 - Material number: 1044854, Catalog number B1018-284.
Distribution pattern: Worldwide Distribution-USA (nationwide) and the countries of Algeria Argentina Brazil Canada China Chile Columbia France Indonesia Libya Georgia Germany India Italy Japan Kazakhstan Lithuania Oman Pakistan Peru Poland Portugal Republic of Korea Romania Russia Saudi Arabia Spain Turkey.
Derived failure modes
-
Unknown
Springs contained in the access door hinge assembly on the Walk Away Plus instruments are becoming detached causing premature failure of the hinge assembly and a potential safety issue for the customers.