The 5085 is a general surgical table with high patient weight capacity, extended width capability, and lower minimal table top elevation.

Z-0196-2015

October 10, 2014

Class II

Product summary

Firm
Steris Corporation
Event
Event 69472
Status
Terminated
Classification
Class II
Quantity
999
Official record key
device-enforcement:Z-0196-2015

Official wording

Reason: The 5085 and 5085SRT stainless steel column shrouds may become damaged as a result of misalignment during extreme table articulation, excessive side force on the table column, or pressure on the base of the table. Severe damage to the shrouds may prevent the user from raising or lowering the surgical table.

Code information: 5085 and 5085SRT

Distribution pattern: US distribution: AL, AZ, AR, CA, CO, CN, DC, FL, GA, HI, ID, IL, IN, IA, KN, KY, LA, ME, MD, MI, MN, MS, MO, NB, NV, NY, NC, ND, OH, OK, OR, PN, RI, SC, SD, TN, TX, VA, WV, WI. Foriegn: Austrialia, Brazil, Canada, China, Ecuador, India, Korea, Mexico, Pnama, and Thailand.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The 5085 and 5085SRT stainless steel column shrouds may become damaged as a result of misalignment during extreme table articulation, excessive side force on the table column, or pressure on the base of the table. Severe damage to the shrouds may prevent the user from raising or lowering the surgical table.