Skip to content
Recall Observatory FDA recall evidence

Device product

Nova StatStrip/StatSensor Hospital Blood Glucose Meter; Nova StatStrip Lactate Hospital Meter System; Nova StatSensor Creatinine Hospital Meter, Rechargeable Lithium Batteries Single: 46626; 4 Pack: 46827; 5 Pack: 50436. For Point-Of-Care usage in the quantitative determination of Glucose in whole blood.

Z-0650-2014

October 10, 2013

Class II

Product summary

Firm
Nova Biomedical Corporation
Event
Event 66398
Status
Terminated
Classification
Class II
Quantity
est. 200,000
Official record key
device-enforcement:Z-0650-2014

Official wording

Reason: Revised Battery Instructions for Removal and Expiration Date due to battery swelling and cell leakage

Code information: All serial numbers

Distribution pattern: Nationwide Distribution.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Revised Battery Instructions for Removal and Expiration Date due to battery swelling and cell leakage