Device product
PT Link Waterbath, Model#PT100. Lab Vision PT Module Plus provides an automated standarized alternative to labor intensive and time consuming dewaxing and epitote retrieval procedures.
Z-0287-2013
Product summary
- Event
- Event 53336
- Status
- Terminated
- Classification
- Class II
- Quantity
- 136 units
- Official record key
device-enforcement:Z-0287-2013
Official wording
Reason: Dako is recalling the PT Link Waterbath because of a possible electrical hazard. Specifically, if the heater pad under the reagent tank becomes wet, it may present an electrical shock hazard when the lid is open.
Code information: Catalog #: PT-Module
Distribution pattern: Nationwide distribution: USA including states of: AK, CA, CO, CT, DC, FL, GA, HI, IA, IL IN, KS, KY, MA, MD, MI, MN, MO, NE, NJ, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, and WA.
Derived failure modes
-
Unknown
Dako is recalling the PT Link Waterbath because of a possible electrical hazard. Specifically, if the heater pad under the reagent tank becomes wet, it may present an electrical shock hazard when the lid is open.