Skip to content
Recall Observatory FDA recall evidence

Device product

Site~Rite 8 Ultrasound Systems-intended for diagnostic ultrasound imaging of the human body. Site~Rite 8 Ultrasound System, Catalog No.9770500 Refurbished, Site~Rite¿ 8 System, Catalog No.9770500R Site~Rite 8 Ultrasound System, Catalog No.9770501 Refurbished Site~Rite¿ 8 System,Catalog No.9770501R Site~Rite 8 Ultrasound System, Catalog No. 9770550 Site~Rite 8 Ultrasound System, Catalog No. 9770552 Site~Rite 8 Ultrasound System, Catalog No. 9770553 Site~Rite 8 Ultrasound System, Catalog No. 9770554 Site~Rite 8 Ultrasound System, Catalog No. 9770555 Site~Rite 8 Ultrasound System Console with Connector, Catalog No.: 9770600 Site~Rite 8 Ultrasound System Console with Connector, Refurbished, Catalog No: 9770600X Site~Rite 8 Ultrasound System Console with Connector (Remanufactured) Catalog No: 9770600Y

Z-0518-2020

November 04, 2019

Class II

Product summary

Firm
Becton Dickinson & Company
Event
Event 84262
Status
Terminated
Classification
Class II
Quantity
2215 units
Official record key
device-enforcement:Z-0518-2020

Official wording

Reason: The indicator may show available charge when the battery is close to depletion. As a result, the Site~Rite 8 has the potential to abruptly shut down, without advance warning to the user, when running on battery power only

Code information: All Serial numbers

Distribution pattern: Nationwide Foreign: UK, Belgium, Australia, Japan, Canada, Taiwan, Colombia, Chile, Philippines, Thailand, Brazil, Hong Kong, Chile, Argentina, Mexico, India, Singapore

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The indicator may show available charge when the battery is close to depletion. As a result, the Site~Rite 8 has the potential to abruptly shut down, without advance warning to the user, when running on battery power only