pO2 D999 membrane units. Distributed by Radiometer America, Inc., Westlake, OH 44145 pO2 membrane units, ordering number 942-042 Intended use is in vitro testing of samples of whole blood for pO2 (partial pressure of oxygen).

Z-0713-2014

June 10, 2013

Class II

Product summary

Firm
Radiometer America Inc
Event
Event 67044
Status
Terminated
Classification
Class II
Quantity
209 boxes/6 membranes
Official record key
device-enforcement:Z-0713-2014

Official wording

Reason: RADIOMETER became aware that the regarded lot pO2 units 942-042R0971 were produced with inner solution for pCO2 membrane units. This solution is a salt solution similar to the pO2 inner solution, but also contains glycerol which can be reduced at the cathode causing a suddenly rising zero current, causing the membrane unit to fail the following Cal2 calibration.

Code information: Part #: 942-042, Lot #: R0971

Distribution pattern: Worldwide Distribution - US (nationwide) including states of: CO, CT, GA, MA, MO & MS and country of: Canada.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    RADIOMETER became aware that the regarded lot pO2 units 942-042R0971 were produced with inner solution for pCO2 membrane units. This solution is a salt solution similar to the pO2 inner solution, but also contains glycerol which can be reduced at the cathode causing a suddenly rising zero current, causing the membrane unit to fail the following Cal2 calibration.