Skip to content
Recall Observatory FDA recall evidence

Device product

Heater Cooler Unit Usage: The Heater-Cooler Unit (HCU 30) supplies temperature-controlled water for cardioplegia heat exchangers and for blood heat exchangers in extra corporeal circulation and for blankets with which patients can be warmed or cooled. The HCU 30 is intended to circulate water through heat exchange circuits to warm or cool a patient during short duration cardiopulmonary bypass procedures lasting 6 hours or less.

Z-0348-2017

September 27, 2016

Class II

Product summary

Firm
Maquet Cardiovascular Us Sales, Llc
Event
Event 75271
Status
Terminated
Classification
Class II
Quantity
305 units
Official record key
device-enforcement:Z-0348-2017

Official wording

Reason: There is the potential for contaminated water to enter parts of the Heater-Cooler Unit circuits. There is a potential risk of bacterial contamination including Mycobacteria species.

Code information: HCU 30 Base Unit 100-120 V, Model Number 704629 I 70103.4371 HCU 30 Base Unit Canadian/US, Model Number 705501 I 70103.4653 HCU 30 Base Unit 200-240 V, Model Number 939001 I 70102.8718 All serial numbers are affected.

Distribution pattern: Worldwide Distribution - US Nationwide

Derived failure modes

  • Microbial contamination

    reason.microbial_contamination · v1.0.0

    bacterial contamination