Device product
Siemens ADVIA 1200 and ADVIA 1650 Chemistry Systems. Automated, clinical chemistry analyzers that can run tests on human serum, plasma, or urine.
Z-0800-2014
Product summary
- Event
- Event 67030
- Status
- Terminated
- Classification
- Class II
- Quantity
- 1563
- Official record key
device-enforcement:Z-0800-2014
Official wording
Reason: An overflow flag //// is displayed in the result column for a test when the algorithm for the test cannot calculate a result or when the concentration cannot be determined because it is outside the absorbance limits, and it is possible for the user to misinterpret or edit the result.
Code information: ADVIA 1200 with software versions: V2.00 and V2.01 ADVIA 1650 with software versions V3.52 and V4.01
Distribution pattern: Worldwide Distribution. USA nationwide and the country of Canada.