Device product
BD Alaris Infusion Pump Module Model 8100 Pump Module Keypad; and Pump Module Door Assembly Replacement Kit with Pump Module Keypad, Parts: 49000239; 49000346; 49000438; 49000439.
Z-1309-2021
Product summary
- Event
- Event 87387
- Status
- Ongoing
- Classification
- Class I
- Quantity
- 160300
- Official record key
device-enforcement:Z-1309-2021
Official wording
Reason: Pump Module keypad lifting, and Fluid ingress could result in 1) Unresponsive keys: module continues infusion, PC unit will not alarm and must be used for programming changes, may necessitate a different Pump Module. May result in infusion start delay/inability to titrate medication. 2) Stuck keys: PC unit alarms, module exhibits Channel Error, may result in infusion interruption/start delay
Code information: Pump Module with Keypad: manufactured January 23, 2019 to December 5, 2019. Pump Module Door Assembly Kit with Keypad: Dated from January 15, 2019 to November 14, 2019.
Distribution pattern: Worldwide distribution. US nationwide and countries of: AE, IL, IN, SA, BE, QA, SG, TW, KW, ZA, PH, AU, HU, NZ
Derived failure modes
-
Unknown
Pump Module keypad lifting, and Fluid ingress could result in 1) Unresponsive keys: module continues infusion, PC unit will not alarm and must be used for programming changes, may necessitate a different Pump Module. May result in infusion start delay/inability to titrate medication. 2) Stuck keys: PC unit alarms, module exhibits Channel Error, may result in infusion interruption/start delay